High-Power Laser DIOLAS LFD Advanced

DIOLAS LFD Advanced is a powerful laser system which can effectively coagulate and cut tissue. Different to normal coagulation lasers, it is able to limit the effects in deeper tissue levels, depending on the operating mode.

Surgical side-effects such as swelling, inflammation or organ dysfunction are strongly reduced while the defined, constant coagulation closes even larger blood vessels.

This powerful laser device is the perfect interdisciplinary tool for surgeons in most medical fields that require strong and controlled coagulation.

Main Features

  • Ability to selectively enable a deep-level or more superficial coagulation, depending on the program selected
  • Large 10,4 " True Color Touch Panel for easy and intuitive use
  • Low noise level and highly mobile by large castor wheels
  • Strongly recommended for interdisciplinary use

  • image
  • image
Previous Next

Brochure DIOLAS LFD Advanced

Treatment Pictures & Videos

Please click on the video's fullscreen button to enlarge.

Medical Applications


Please click on the individual disciplines to see the type of treatments this laser device can be used for.

Available Configurations

  • DIOLAS LFD Advanced

Main Accessories

Technical Specifications

Laser type Semiconductor-based Laser module with LFD (Levelled Field Density) Extension
Operating modes & power at tissue
  • LFD Mode
  • Continuous Wave
  • Pulsed Mode (milliseconds or microseconds range)

Output power freely adjustable between 1 - 200 Watt

Activation profiles
  • Continuous
  • Pulse repetition
User interaction
  • True Color Touch Panel 10,4 Inches (26,4 cm) with brilliant colors
  • Adjustable audio output on emission and ready / standby
  • USB Port for output of treatment protocols
  • Output power can be controlled by footswitch in 5 Watt steps
  • Integrated treatment protocol with USB export and individual patient ID
Beam delivery Various types of flexible fiber probes
Pilot beam

red (635 nm), < 3 mW, adjustable

Integrated ports
  • USB port
  • Control of external smoke evacuator VACULAS
  • Warning lamp system
Memory for user programs Storage for up to 9 individual operating parameters
Cooling Internal
Electrical
requirements
230 V (10 A) / 110 V (16 A), 50 / 60 Hz, single-phase
Dimensions & weight 50 x 100 x 95 cm (W x H x D), 102 kg

Technical changes reserved.