Micromanipulators Print

Micromanipulator MicroSpot


Colour-Corrected Micromanipulator MICROSPOT ZOOM

  • Adjustable focal length between 200 - 400 mm.
  • Focal spots of CO2 and HeNe laser beams congruent by new mirror technology, no lenses!
  • 85% of incoming laser energy is being transmitted.
  • The minimal focal spotsize of the pilot beam of 0,15 mm in combination with a focal length of 400 mm produces a minimal diffusion on the tissue surface and therefore allows a better sight on it.

Laser beam diameter in dependance of the working distance:

Working distance Min. focused Max. defocused
f = 200 mm 0,100 mm 2,5 mm
f = 250 mm 0,115 mm 3,0 mm
f = 300 mm 0,125 mm 3,5 mm
f = 350 mm 0,130 mm 4,0 mm
f = 400 mm 0,150 mm 4,5 mm
To be mounted with dovetale adapter.
Micromanipulator Microspot

In Particular Interesting for the Disciplines

  • Neurosurgery
  • ENT
  • Gynaecology

Micromanipulator MiniSpot


Some Particularities:

  • Focal length 200 – 400 mm
  • Max. defocusing 12x
  • Hand support can be rotated between 0 – 220 degrees
  • Connectable to laser from left or right side by adapter
  • To be mounted with dovetale adapter
  • Min. focus aperture:
    • 0,32 mm at 200 mm focal length
    • 0,45 mm at 250 mm focal length
    • 0,53 mm at 300 mm focal length
    • 0,74 mm at 400 mm focal length
Micromanipulator MiniSpot

In Particular Interesting for the Disciplines

  • ENT
  • Gynaecology



Related Links

Compatible with Laser



Technical changes reserved